You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
The rear detector strip beam stop moves left-right (x-axis) across the face of the front detector array.
The front detector strip beam stop is not often used (usually only during commissioning). The control could be located on an Advanced tab (possibly Manager password protected).
Because the motors that move the beam stops operate in a vacuum, they should be de-energised when a move is complete (otherwise they are at risk of over-heating).
There are schematic diagrams showing the operation of the SANS2D vacuum tank.
There are schematic diagrams showing the how the SANS2D vacuum tank User OPI and Advanced OPI could look.
The LabVIEW VI for the SANS2D vacuum tank is located at: C:\LabVIEW Modules\Instruments\SANS2D\Sans2d Tank (for more details see the S2d Tank - Front Panel.vi in the file Sans2d Tank - System Functions.llb)
Check the paragraph describing the Strip Beamstops in the Beam_line_descriptors document.
As a SANS2D scientist, I want IBEX to provide control for the front detector strip beam stop, so that I can control its position.
Acceptance Criteria
Notes
C:\LabVIEW Modules\Instruments\SANS2D\Sans2d Tank(for more details see theS2d Tank - Front Panel.viin the fileSans2d Tank - System Functions.llb)